Calectro

Portable Hand Held CO2 Meter - Product Specifications

CO 2 Measurement:

Operating PrincipleNon-dispersive infrared (NDIR) with gold plated optical cell
Gas Sampling Mode Diffusion
Response Time (1/e) 2 min diffusion time & 15 sec at 0.2 litre/min gas flow
Measurement Range 0-3 % vol.
Extended Range 3-10 % vol. (accuracy not specified)
Accuracy at NTP (+25° C) ± 3 % of reading or ± 0.02 % vol., whichever is greater
Pressure Dependence + 1.6% reading increase per kPa deviation from normal pressure
Temperature Dependence =0.005 % vol. / °C at zero gas level
=0.015 % vol. / °C at 3 % vol. CO 2
Time Weighted Average (TWA) calculation 8 h time span (most resent) with 4 min sample period
Reset can be selected during unit turn-on sequence
Sleep Mode "Over-night" charging mode with activated ABC algorithm

Alarm / Measurement Interface:

LEDs 5 step "bar graph" green-green-yellow-yellow-red LEDs with trip points defined by the present CO 2 concentration and preset comparator levels
Numerical Liquid Crystal Display Simultaneous display of
* the current CO 2 concentration (in % vol.)
* the 8 h CO 2 TWA value (in ppm)
* battery status indication
Audiable horn 80 dBR/10cm transducer with 2kHz resonance frequency, sounding during alarm status until push-button acknowledgement is pressed
Push-button Alarm status acknowledgement and back-light on/off switch
Internal Data Logger Event triggered data recordings with real clock time stamps, plus the latest 8 h accumulated turn-on time of recorded CO 2 concentration data.
Digital Interface USB cable with sensor UART-RS232 com driver

Electrical:

Battery Charger Input 6 VDC / 500 mAh, with NOKIA type miniature connector
Internal Battery 3,6 VDC / 1400 mAh Li-ion accumulator ( > 8 h. capacity)
Current Consumption< 55 mA @ in normal mode
< 100 mA @ in alarm mode

General Performance:

Compliance with EMC Directive 89/336/EEC
Storage Temperature Range -20° to +70° C
Operating Temperature Range 0° to +50° C
Operating Humidity Range 0 to 95 % RH (non Condensing)
Sensor Life Expectancy > 15 years
Battery Life Expectancy > 3 years
Self-diagnostics complete power/sensor/ internal checks
Status Indicator LCD triangle ikon = maintenance call
Power-up Time < 30 sec. (full specs < 15 minutes)
Housing Material ABS/PC blend
Dimensions (L x W x D) 125 x 50 x 35 mm
Total Weight 120 g

Click here to view/download pdf file